Preparação do meio Série MediPrep Dados Técnicos
| Weight | 36 kg (MediPrep 820)// 41 kg (MediPrep 1622) |
|---|---|
| Dimensions (H x W x D) | 675 x 354 x 635 mm (MediPrep 820)// 692 x 367 x 645 mm (MediPrep 1622) |
| Voltage | 115 V or 230 V, 50/60 Hz |
| Interfaces | USB-A for printer, RS-232 interface for firmware |
| Fuses | 2x15 A |
| Max. volume inside tank | 8 liters (MediPrep 820)// 16 litres (MediPrep 1622) |
| Min. volume inside tank | 2 liters |
| Functions | Vacuum degassing, heating, gravimetric filling (MediPrep 820)// Mixing, diluting, vacuum degassing, heating, gravimetric filling (MediPrep 1622) |
| Pharmacopeia conformance | USP, EP |
| Inlet for premixed media | 1 Inlet for media from external source (MediPrep 820)// 1 Inlet for media from external source, 2 inlets for concentrates (MediPrep 1622) |
| Outlet for residual or waste medium | yes |
| Remote dosing hand | Optional |
| Control | Touchpanel 7“ |
| Common aqueous media at least 2 l | yes |
| Time needed for media preparation | ≤ 15 minutes for 8 l, depending on the degassing time (MediPrep 820)// ≤ 24 minutes for 16 l, depending on the degassing time (MediPrep 1622) |
| Mixing accuracy | Mixing not possible (MediPrep 820)// ≤ 1 % with a ratio of concentrate to water 1: 9 to 1: 1 (m / m) (MediPrep 1622) |
| Handling of foaming media (SDS) | Up to 1 % solution at max. 6 l (MediPrep 820)// Up to 1 % solution at max. 14 l (MediPrep 1622) |
| Dosing volume, accuracy | ≤1% for 250 to 1000 ml (MediPrep 820)// ≤1% for 500 to 1000 ml (MediPrep 1622) |
| Weighing accuracy | ≤ 0,1 % at 2 kg |
| Heating temperature, accuracy | Max. 50 ° C, ± 0.2 ° C |
| Degassing pressure, accuracy | ≤ 200 mbar, ±10 mbar |
| Product memory | Max. 100 |
| User memory | Max. 50 users, password-protected access |